Home
 
You are here: Products > Pressure > Handhelds > PH720
 

Products

 

PH720

The model PH720 electronic balancing tool is auto-zeroing and is density corrected for temperature and pressure. Users can power the unit by AC or with 4 AA-size rechargeable or alkaline batteries. It comes with a pitot probe and 2 static pressure probes, but it can be used with available 16-point velocity matrix, air foil, temperature, or RH/temperature probes. CompuDAT™ PC downloading software is included.

Replaces:
EBT720 (Alnor)

  PH720
 
Specifications AXD
Range
diff. pressure ±15 in. H2O, (3735 Pa) 150 in. H2O maximum safe operating pressure
absolute pressure 15 to 40 in. Hg (356 to 1016 Hg)
velocity 0.125 to 40 m/s pitot probes;
0.125 to 25 m/s air flow probe;
0.125 to 12.5 m/s velocity matrix;
volume 25 to 2,500 ft3/m (42 to 4250 m3/h) capture hood
RH 0 to 95% RH (optional probe)
temperature -40 to 121°C probe dependant
Resolution
pressure 0.00001 in. H2O (0.001 Pa) Static & Differential Pressure
0.001 in. Hg (1 mm Hg) Absolute Pressure
velocity 0.1 ft/min (0.1 m/s)
volume 0.1 ft3/min (0.1 m3/h)
RH 0.1% RH
temperature 0.1 °C
Accuracy
pressure ±2% of reading ±0.025 mm H2O; ±2% of reading ±0.001 in. Hg
velocity

±3% of reading ±0.04 m/s25 to 8,000 ft/min (all velocity probes) > 50 ft/min

volume ±3% of reading ±7 ft3 /min 25 to 2,500 ft3/min > 50 ft/min
RH ±3% RH
temperature ±0.3°C from 0 to 71 °C typically ±0.6°C from -40 to 0°C and from 71 to 121 °C

Applications
Pressure measurement
Velocity measurement
Volume measurement
 
© ACIN instrumenten bv - handelskade 76 - 2288 bg rijswijk - the netherlands - info@acin.nl