Calibration
devices
The Rotronic calibration devices are small, airtight chambers that
fit exactly to the respective probes. The lower part is a screw-mount
lid, into which the humidity standard is filled, together with a textile
pad. After a period of stabilisation, typically 60 minutes at stable
temperature, the required humidity value is attained. The instrument
can now be calibrated against the humidity reference value. Calibration
devices are available for the entire range of Rotronic products, and
we can also supply calibration devices for third party probes.
The inner volume of the calibration device should not be too big;
it should not exceed 50 ml of free space when the probe is mounted.
When the volume is too big, the correct humidity may eventually
not be generated and the calibration will be wrong. |
|
 |
| |
Specifications Calibration
devices
| EM-G |
screw-in probe type E, HP...E... screw in |
| EGS |
for all ROTRONIC sword probes push-on |
| EAS |
for bell probe AS-ST |
| EAWC |
for AWC-measuring station |
| EBFC |
for web probe BFC |
| ERA |
for bell probe for rolls |
| WP-14 |
for probe AWD, AWVC |
| ER-15 |
probe 15 mm ø push-on type |
| ER-05 |
probes with 4/5 mm ø push-on |
| EGL |
probe 10 mm ø (A1, M1) push-on |
| ER-18K |
probe 18 mm ø push-on |
| ER-20K |
probe 20 mm ø push-on |
| ERV-15 |
probe 15 mm ø push-on / vertical |
| EM-15 |
probe 15 mm ø screw-in |
| EM-25 |
probe 25 mm ø screw-in |
| EMV-25 |
probe 25 mm ø screw-in / vertical |
| EMV-15 |
probe 15 mm ø screw-in / vertical |
| EDM 15/25 |
2 probes, 15 & 25 mm ø screw-in |
| Applications |
| Calibration on-site, in workshops on laboratories |
|